When performing an ATP test, it's crucial to adhere to safety guidelines to protect yourself and others. Refer to the Safety Data Sheets (SDS) or all safety guidelines related to our reagents, and follow all safety protocols at your testing location.
Gloves are recommended when handling wastewater and chemical product samples. However, caution should be exercised when using gloves during the ATP assay (the final step). Static buildup and discharge between the gloves and tubes can interfere with the assay and cause high backgrounds. To avoid this, insert the assay tube into the luminometer with an ungloved hand.
LuminUltra's enzymes require special handling and storage conditions to maintain optimal activity. All other reagents are safe to handle, but consult the Test Kit Instructions and SDS for recommendations for storage and handling.
For further assistance or questions, please contact Customer Experience by emailing customerservice@luminultra.com or filling out a customer support ticket at the following link: Customer Support Ticket.